Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8600A: (1 year) CAL VER /5720,5725 DATE: 01-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 42 NUMBER OF LINES: 120 CONFIGURATION: Fluke 5720A CONFIGURATION: Fluke 5725A ============================================================================= # # Source: # Fluke 8600A manual # # Compatibility: # MET/CAL 4.1 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5720A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5720A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F W 1.002 ASK+ K 1.003 M5720 C2 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Before performing verification allow the UUT to 1.005 DISP stabalize (30 min.) @ 23+/-5degC. 1.005 DISP Humidity less than 80% 1.006 DISP Turn on UUT. 1.007 HEAD {DC VOLTAGE TESTS} 1.008 DISP Connect the 5720A and the UUT as follows. 1.008 DISP [32] 5720A OUTPUT HI to UUT V-Ohm HI 1.008 DISP [32] 5720A OUTPUT LO to UUT V-Ohm LO 1.008 DISP [32] 5720A SENSE HI to UUT V-Ohm HI 1.008 DISP [32] 5720A SENSE LO to UUT V-Ohm LO 1.008 DISP [32] 5720A AUX CURRENT OUTPUT HI to UUT mA 1.009 HEAD DC VOLTAGE TESTS: {200mV Range} 1.010 5720 200 190.00mV 0.1U 2W 2.001 5720 200 -19.00mV 0.03U 2W 3.001 HEAD DC VOLTAGE TESTS: {2V Range} 3.002 5720 2 1.9000V 0.0005U 2W 4.001 5720 2 -1.9000V 0.0005U 2W 5.001 5720 2 0.9000V 0.0003U 2W 6.001 HEAD DC VOLTAGE TESTS: {20V Range} 6.002 5720 20 19.000V 0.005U 2W 7.001 5720 20 -19.000V 0.005U 2W 8.001 HEAD DC VOLTAGE TESTS: {200V Range} 8.002 5720 200 190.00V 0.05U 2W 9.001 HEAD DC VOLTAGE TESTS: {1200V Range} 9.002 5720 1200 1000.0V 0.3U 2W 10.001 HEAD DC VOLTAGE TESTS: {Autorange} 10.002 RNG V 10.003 5720 -170.00mV 0.09U 2W 11.001 5720 -1.7000V 0.0004U 2W 12.001 5720 -170.00V 0.04U 2W 13.001 5720 -1000.0V 0.3U 2W 14.001 HEAD {RESISTANCE TESTS} 14.002 HEAD RESISTANCE TESTS: {Autorange} 14.003 RNG kZ 14.004 5720 0.0000Z 0.03U CW 15.001 5720 100.00Z 0.13U CW 16.001 5720 1.0000kZ 0.0011U CW 17.001 5720 10.000kZ 0.006U CW 18.001 5720 100.00kZ 0.06U 2W 19.001 5720 1000.0kZ 0.6U 2W 20.001 5720 10.00MZ 0.021U 2W 21.001 DISP Remove the connections to the UUT V-Ohm HI terminal. 21.002 EVAL Is the display flashing and the 20 MOhm LED ON? 22.001 DISP Reconnect the OUTPUT HI and SENSE HI terminals of 22.001 DISP the 5720A to the UUT V-Ohm HI terminal. 22.002 HEAD {DC CURRENT TESTS} 22.003 RNG 200uA 22.004 MATH MEM1 = 0.0 22.005 MEMI Enter UUT reading in microamps: 22.006 MEME 22.007 MEMC 200 uA 0.05U 23.001 5720 2 1.9000mA 0.0021U 2W 24.001 5720 20 19.000mA 0.021U 2W 25.001 5720 200 190.00mA 0.21U 2W 26.001 5720 2000 1900.0mA 2.1U 2W 27.001 HEAD {AC VOLTAGE TESTS} 27.002 HEAD AC VOLTAGE TESTS: {200mV Range} 27.003 5720 200 190.00mV 1.95U 70kH 2W 28.001 5720 200 190.00mV 0.54U 500H 2W 29.001 5720 200 39.00mV 0.4U 30kH 2W 30.001 HEAD AC VOLTAGE TESTS: {2V Range} 30.002 RNG 2V AC 30.003 5720 0.0000V S 2W 30.004 MEMI Enter UUT reading in volts AC: 30.005 MEME 30.006 MEMC 2 V 0.0040U 31.001 5720 2 1.9000V 0.0041U 500H 2W 32.001 5720 2 0.9000V 0.0021U 500H 2W 33.001 5720 2 1.9000V 0.010U 50kH 2W 34.001 5720 2 1.9000V 0.02U 100kH 2W 35.001 5720 2 1.0000V 0.0055U 30H 2W 36.001 HEAD AC VOLTAGE TESTS: {20V Range} 36.002 5720 20 19.000V 0.041U 500H 2W 37.001 5720 20 19.000V 0.2U 70kH 2W 38.001 HEAD AC VOLTAGE TESTS: {200V Range} 38.002 5720 200 190.00V 0.41U 500H 2W 39.001 5720 200 100.00V 0.55U 30kH 2W 40.001 HEAD AC VOLTAGE TESTS: {1200V Range} 40.002 5720 1200 1000.0V 3.1U 500H 2W 41.001 5720 1200 1000.0V 6.0U 20kH B1 2W 42.001 END